A spring-loaded test probe looks unchanged right up until it starts failing intermittently, which is exactly why it gets little attention. The failure is gradual: a few million actuations, a thin gold layer worn through, a barrel interface that no longer seats cleanly. By the time a fixture starts throwing first-pass failures, the probes have been drifting for weeks.
Gold is there to break through oxide and keep the contact stable, not to last forever. On a pointed or crowned tip the gold is thin, often a flash over nickel, and the tip takes the most rubbing. Once it is through, the underlying nickel or brass oxidises faster than the original surface, and contact resistance climbs with humidity rather than staying flat. A probe that read 40 mΩ at commissioning and now reads 90 mΩ is not "still fine"; it is on its way out.
| Symptom | What it usually means |
|---|---|
| Resistance creeps up over weeks | Gold worn through, base metal oxidising |
| Intermittent opens that clear on retest | Plunger sticking or barrel losing seat |
| High force needed to get a reading | Burnt or deformed tip, or contaminated barrel |
| Scatter across a board, not one net | A set of mixed old and new pins |
The plunger slides inside the barrel on a second contact that people forget about. It adds its own resistance in series with the tip, and it wears too. A probe can have a perfect tip and still read high because the plunger no longer seats cleanly in the barrel. You cannot see this from outside, which is why it shows up as mystery drift.
A tip that has gone flat, picked up a bright polished band, or lost its crown shape is past its life regardless of the number on the meter. A plunger that does not return smoothly when depressed by hand is sticking and will give intermittent opens. Neither shows up in a photograph of the fixture from above, so the check has to be hands-on, probe by probe, on a schedule.
Decide the end-of-life rule before you need it: a resistance above a set multiple of baseline, a visible tip change, or a count of actuations, whichever comes first. Cleaning helps a contaminated tip but cannot restore worn gold; once the plating is gone, replacement is the only fix. A planned pin swap during a maintenance window costs far less than a shift of unexplained first-pass failures.