Every spring probe datasheet quotes a force and a contact resistance, and the two numbers are usually printed in different columns with no relationship stated. In use they are tightly coupled: the force sets the real contact area, the real contact area sets the resistance, and the resistance sets whether a 30 mOhm limit means anything on a 20 mOhm part.
Measured resistance across a mated contact has three parts, and only the first is a property of the metal:
Constriction resistance falls roughly as the inverse square root of contact force, because the contact area grows with the deformation. That is the physics behind the curve everyone knows informally: doubling force does not halve resistance.
| Applied force | Typical contact resistance | What is happening |
| 0.3 N | 100 to 200 mOhm | Few asperities in contact, films intact |
| 0.5 N | 60 to 120 mOhm | Contact area growing quickly with force |
| 1 N | 30 to 60 mOhm | Plateau begins; films broken at asperity peaks |
| 2 N | 20 to 40 mOhm | Little gain; wear rate rising |
| 4 N | 15 to 30 mOhm | Plateau; plating damage becomes the limit |
The bands are typical rather than guaranteed, but the shape is reliable: between 1 N and 2 N most probes have already reached the part of the curve where force buys very little resistance. Going to 4 N buys a few more milliohms and costs a great deal of plating life.
Spring probes are specified at a force measured at a stated deflection, usually two thirds of rated travel. A probe rated 1.2 N at 2/3 travel delivers roughly 0.8 N at 1/3 travel and about 1.5 N at full travel, because the spring is close to linear over its working range.
This means the force in a fixture is not the datasheet number unless the fixture plunges to the stated point. Two fixtures built to the same drawing can differ by 30 percent in contact force if one plunges 1 mm and the other 3 mm. The symptom is a resistance reading that is stable on one station and marginal on another, with the same probe part number.
A probe that reads 25 mOhm at 08:00 and 60 mOhm at 16:00 has not changed its spring. The usual causes, in order:
The second and third are cumulative and irreversible. This is why a 60 mOhm reading should be treated as a wear indicator, not a calibration problem.
IEC 60512-2-1 defines the four-wire method for connector contact resistance, and it is the only defensible method for values in this range. A two-wire measurement adds the lead and contact resistance of the meter's own probes, which can easily be 30 to 80 mOhm — the same order as the value being measured. With four wires, the sense leads carry no current and the lead resistance cancels.
Two further rules keep the number honest:
The test probe and test pin range lists probes by rated force at 2/3 travel; the catalogue indexes the barrels and receptacles that set the mechanical side of the same interface.