An accessibility test ends in a yes or no: did the probe touch a live part. The probe supplies geometry; the indicator supplies the answer. Most of the argument at the end of a test is about the indicator.
The circuit is usually a low-voltage lamp in series with the probe tip and the supply, with a resistor to limit current. The voltage is typically 40 V to 50 V, the current a few milliamps, and the lamp is chosen so that a film of contamination does not light it.
The indicator must not damage the sample. At 40 V a thin insulating film survives the test, which is the point: a real user touching the same surface would not break down a 0.05 mm film either. Testing at mains voltage would break the film and report a hazard that does not exist in service.
| Element | Typical value | Effect on the verdict |
| Indicator voltage | 40 V to 50 V AC or DC | High enough to light a lamp, low enough not to puncture films |
| Series resistor | Sets current to a few mA | Protects the sample and the lamp |
| Lamp | Filament or LED with threshold | Filament shows short contact; LED may miss it |
| Probe tip material | Brass or steel, clean | Oxide on the tip reads as an insulator |
| Contact area | Tip radius, not the whole finger | A 2 mm radius is the whole contact patch |
| Circuit return | To the earthed enclosure | A floating return gives no reading |
A filament lamp is slower than an LED and that is an advantage: it averages out intermittent contact and shows a steady glow only where the contact is real.
For every access point, record the probe used, the force, the result, and the depth of insertion if the probe could move. A report that says "no contact" for twelve openings tells the designer nothing about which opening was 1 mm from contact.
The indicator circuit is part of the instrument, not an accessory: it should be checked against a known live part at the start of every session, which takes ten seconds and catches the dirty tip, the broken lead and the flat battery in one go. Our probe sets include the lead and lamp holder, and the geometry of each probe is listed in the technical notes.
Standards referenced: IEC 61032, IEC 60529 clause 13.1, IEC 60950-1 clause 2.1.1 (indicator circuit practice), UL 507.
An accessibility test that passes a 4 mm gap reports a design as safe when a 12 mm finger reaches a live terminal at 230 V. The cost of that error is not in the laboratory; it is in the field, in a recall that starts with a single incident and ends with a 5,000 unit return.