Home > Technical Articles
2026-11-19
The probe touches or it does not. In theory the detection circuit is trivial. In practice the circuit is a voltage source, a series resistor, a lamp or a detector, and two contacts whose area may be a fraction of a square millimetre. Every element contributes to whether a touch is reported.
The detection circuit has to register a genuine contact through the tiny contact area of a probe tip, without applying a voltage high enough to be a hazard itself, and without drawing enough current to weld or mark the surface being tested.
| Element | Typical value | What it decides |
| Source voltage | 40 V | Enough to break down thin films, low enough to be safe |
| Series resistor | 220 kΩ to 1 MΩ | Limits current to a fraction of a milliamp |
| Lamp or detector | Neon, LED or relay | Threshold and response time |
| Contact area at the tip | Under 1 mm² | Whether a light touch registers |
| Surface film | Oxide, paint, flux | The most common cause of a false pass |
Most false passes are not electrical, they are mechanical. The probe tip stops 0.1 mm short, the lamp stays off, and the report says not accessible. A 0.1 mm gap is 4 thousandths of an inch, well inside the tolerance of most moulded parts, and it is exactly the gap a thin oxide film will bridge or a slightly worn tip will produce.
The second source is the resistor. A 1 MΩ series resistor with a neon lamp draws so little current that a high-resistance path through a damp surface can look like contact in one direction and nothing in the other. Using a detector with a defined threshold, rather than a lamp that glows a little, removes the ambiguity.
Thirty seconds of checks, once per session, is the cheapest protection against a report that says not accessible when the truth was not detected.
Standards referenced: IEC 60529 clause 13 for the access conditions, IEC 61032 for probe geometry, IEC 61010-1 for the safety of the detection circuit itself.
The detection voltage has to be high enough to break down thin films and low enough not to be a hazard in its own right. Thirty to fifty volts is the usual range; 40 V is common because it will break down a light oxide film and still limit current to well under 1 mA through a 220 kΩ resistor.
Raising the voltage to 230 V would break down more films, but it also makes the probe tip a shock hazard if the insulation fails, and it can mark or pit the surface being tested.
At 40 V, a 220 kΩ resistor limits current to 0.18 mA into a short. That is enough to light a neon lamp through its own ballast and not enough to damage a plated contact. A 1 MΩ resistor drops that to 40 µA, which many detectors will not register reliably.
The resistor also protects the operator: if the probe body touches a live part, the current through the handle is limited by the same resistor, not by the source.
A lamp glows gradually as current rises, which makes a marginal contact look like a weak pass. A detector with a defined threshold, such as a comparator or a relay, gives a yes or no at a stated current, which is what a test report needs.
Whatever is used, verify it against a known load before each session: a 100 kΩ resistor between probe and source should light the lamp, a 10 MΩ resistor should not.