Home > Technical Articles
2026-11-25
Test probes develop bright bands where they rub. On a probe with a shoulder or a stop face, the band appears at the transition between the working end and the body. It looks like polish. It is actually a record of every test the probe has done, and often an early warning.
Three locations account for most of it:
The shoulder. Where the narrow working section meets the wider body. On probes inserted into openings, this is where the body contacts the enclosure edge, and the contact is a line load rather than a face load.
The stop face rim. The outer edge of the disc that limits insertion depth. It takes the contact with the enclosure surface and rounds over time.
The handle joint. On jointed probes, the pivot area. Wear here shows as play rather than as a bright band, but the two often appear together.
| Location | What wear changes | Direction of error |
|---|---|---|
| shoulder | enters deeper | optimistic (reaches further) |
| stop face rim | tilts more easily | optimistic |
| joint | bends at the wrong angle | either |
The shoulder is a defined diameter at a defined distance from the tip. A probe relies on it to stop at the right depth in a specific class of opening. When the shoulder rounds, the probe slides further in than the drawing allows, and it reaches live parts the standard intended to be out of reach.
The shift is small — a few hundredths of a millimetre per year of heavy use — but it is one-directional. Nothing makes a shoulder grow back.
Not every bright band is wear. Ground surfaces often have a polished appearance from manufacture. Two questions separate them:
Is the band uniform in width? A manufacturing finish follows the geometry; a wear band is widest where contact pressure was highest, usually one side.
Can you feel a step? Run a fingernail across the transition. Wear produces a perceptible step, even a small one. A finish does not.
A probe with a measurable step at the shoulder should be replaced rather than refinished. Refinishing removes material from a dimension that is specified, and a hand-refinished probe is no longer the standard tool. The economic argument for refinishing only holds for large gauges, not for probes.
A polished but still flush band is a monitoring item. Measure the shoulder diameter and its distance from the tip at each inspection, and record the numbers. Two readings a year is enough to see whether the wear is progressing.
Four checks, in order:
Straightness — roll on a flat surface.
Tip condition — magnification, looking for flats and burrs.
Shoulder diameter and position — micrometer and a depth reference.
Stop face flatness — surface plate and a feeler.
A probe that passes all four is a probe whose results can be compared with the previous test. One that passes three is a probe whose results are plausible but not comparable, which is the more dangerous state.