Test Corner or Test Finger: Which One Finds the Live Part First

2026-11-20

Both are fixtures, both are made of metal or plastic, and both appear in the same laboratory. They answer completely different questions. The test corner reproduces the thermal environment of a product installed in a room; the test finger reproduces a hand reaching for a live part. Confusing them is not a rare mistake, because the words sound interchangeable to anyone who has not had to defend a result.

What a test corner is for

A test corner is two vertical panels and a floor, painted matt black, arranged so that a product sits in the corner it would occupy against a wall. It exists because heat loss by radiation and convection depends on what surrounds the product. An appliance tested on an open bench loses more heat than the same appliance pushed against a wall, so its measured temperature rise is lower and the test is optimistic.

The dimension that matters is not the colour but the distance: the product is placed as close to the panels as the installation instructions allow, and if the instructions say 50 mm from a wall, the corner is built to 50 mm. A corner set up at 100 mm because it was easier to clamp measures a different product.

Question being askedFixtureWhat changes if you use the wrong one
How hot does it get in a corner?Test corner, matt black panelsOpen bench under-reports rise by 5 to 15 K
Can a finger reach a live part?Test finger or probe per IEC 61032A corner cannot probe anything
Does the enclosure open under heat?Test corner plus measurementBench testing misses the deformation
Is a small part accessible?Probe 18 or 19 depending on the partA jointed finger is too large to enter
Does a lamp or heater scorch a wall?Test corner at installation distanceBench distance changes the result

What a test finger is for

A test finger is a probe with a defined geometry, a defined joint behaviour and, where relevant, a defined applied force. Its purpose is to determine whether a part is accessible, and its dimensions are not negotiable: the articulated finger is 80 mm long with a 12 mm tip and a 50 mm baffle, and a smaller probe is a different probe with a different scope. Using the articulated finger to check a 5 mm gap tells you nothing, because the finger cannot enter the gap in the first place.

Force matters as much as shape. A probe applied at 10 N and a probe applied at 30 N reach different depths in a compliant gasket, and the two values belong to different standards. Quoting the result without the force is quoting half a test.

The order that avoids rework

Accessibility first, temperature second. If a live part can be touched, the temperature rise result is irrelevant because the product fails earlier in the sequence. Running the thermal test first and discovering an access problem afterwards wastes a day and, worse, tempts a team to argue that the access finding is a formality.

The reverse order also has a cost: measuring accessibility on a cold sample tells you nothing about a housing that softens at 90 °C and opens a gap that was 0.8 mm when cold. The complete sequence is access check cold, temperature rise in the corner, then access check again on the hot sample, with the same probe and the same force.

Setting up a corner that can be defended

  • Panels matt black, at least 20 mm thick, and large enough that the product sees no edge within 200 mm.
  • Distance to each panel taken from the installation instructions, and written on the setup sheet.
  • Ambient thermocouples at 1 m from the corner, shielded from the product, three of them.
  • No draught: a corner in a room with an air-conditioning diffuser overhead will read 3 to 6 K low.
  • Photograph the setup. A corner is only reproducible if the next person can rebuild it.

Standards referenced: IEC 60335-1 clause 11 (heating, test corner arrangement), IEC 61032 (test probes and accessibility), IEC 60529 for probe use in IP testing.